Interleaved samples of CH1 and CH2, when CH2 should be off? #19

Open
mh- opened this issue May 30, 2021 · 1 comment
Open

Interleaved samples of CH1 and CH2, when CH2 should be off? #19

mh- opened this issue May 30, 2021 · 1 comment

Comments

@mh-
Copy link

mh- commented May 30, 2021

Hi Scott,
First of all, thanks for your great work!

I'd like to report a problem when a channel is used for triggering, but not shown on the oscilloscope screen.

Steps to reproduce: I used a DS1104Z (latest FW: 00.04.04.SP4) to sample 2 channels, used CH2 for triggering on the slope of a rectangular signal, then turned off CH2 (the yellow CH2 light is off, CH2 is not shown on the screen).
Then I did a single shot capture, then saved the .wfm file.

Now when I read the file using RigolWFM, CH1 and CH2 are interleaved.
I would have expected to only get CH1.

The effect looks like this:
image
image

The .wfm file is attached:
NewFile1.wfm.zip

Thanks
Michael

@scottprahl
Copy link
Owner

Just saw this report today. Sorry. Unfortunately, there is not much that can be done because I don't know where the triggering information is in the .wfm file. If you reverse engineer something I would be happy to implement it.

A workaround is to save both the trigger and signal channel. The problem is that when the trigger channel is not enabled, there is no way to know that it is in the file or not.