Skip to content
New issue

Have a question about this project? Sign up for a free GitHub account to open an issue and contact its maintainers and the community.

By clicking “Sign up for GitHub”, you agree to our terms of service and privacy statement. We’ll occasionally send you account related emails.

Already on GitHub? Sign in to your account

ADU per photon #44

Open
mpmdean opened this issue Dec 11, 2018 · 1 comment
Open

ADU per photon #44

mpmdean opened this issue Dec 11, 2018 · 1 comment

Comments

@mpmdean
Copy link
Contributor

mpmdean commented Dec 11, 2018

In the centroided data, we need to know how to convert from the ADU provided into real phonons.

@bisogni
Copy link

bisogni commented Feb 21, 2019

From the Fe55 calibration of the high-voltage values required to obtained an effective readout noise of 1e-rms, we extracted the following ADU values:

  CCD Left CCD Right
ADU/e- @1e rms 4.93603161 3.482296963

Example:
Resulting ADU @ 1keV | 1371.11989 | 967.3047119

Sign up for free to join this conversation on GitHub. Already have an account? Sign in to comment