The alarm condition in SRS 104 105 106 107 and 108 does not reference the SLS.
add refenrence to SLS (might be SLS120)
-rename the SLS a per what they are about the text says running breath count when the title says RR.
Remember that requirements should be unambiguous. Only one term should be used. If it is important to note the relation between the two terms then this should be adress in the top section ### ALA5
explain why the SRS that relate to ALA5 do reference the SLS (right now it is not clear for someone who doesnt know what does the running breath count is why it should be monitored)